6.4 General errors
Display LoBat
Instrument does not
react
to keystroke
Display to
toto
to
Probe symbol flashes
Message Sto Ful l
StoFullStoFull
StoFull
Cause Remedy
Batteries almost empty Exchange the batteries (see
section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Hold the <CAL> key
depressed and switch the
instrument on
Cause Remedy
Time-out of the interface Check that the instrument is
connected
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
All memory locations are full Output data storage and clear
data storage