Oxi 3210 Operation
37
ba75795e02 03/2009
z Switch directly to the measured value display with <M>.
z Quit the display and move to the next higher menu level with <F1>/
[Back].
4.5.3 Erasing the measurement data storage
How to erase the measurement data storage is described in section
4.5.2 D
ISPLAYING AND EDITING THE MEASUREMENT DATA STORAGE.
4.6 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for D.O. measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under theReset menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<F1>/[menu] or <ENTER> key.
Setting Default settings
Cal. interval 14 d
Comparison meas. Off
Measured parameter D.O. concentration
Relative slope (S
Rel
)1.00
Salinity (value) 0.0
Salinity (function) Off
Stability control On
Temperature unit °C