ORP voltage ProLab 3000
94
ba75645e03 12/2012
8.2.4 Measurement settings
The following settings are possible for ORP measurements in the Mea-
suring menu:
Menu item Setting Description
Measuring
High resolution () Resolution of the mV display:
() = 0.1 mV
() = 1 mV
Alternative TP () () = Temperature sensor of
another channel is used
Setting of man. temp... -35 ... +150 °C Entry of the manually determined
temperature value. For measure-
ments without temperature sensor
only.
Determine reference... Only if the relative ORP (dU) is
selected as the measured param-
eter
Recorder... Start recorder (see page 147).