ba75646e02 01/2010DRAFT January 28, 2010 2:18 pmProLab 4000 Operating manualpH/ISE/Conductivity measuring instrument with automatic sensor recognition

Strona 2 - © 2010, SI Analytics GmbH

Contents ProLab 400010ba75646e02 01/2010

Strona 3 - DÉCLARATION DE CONFORMITÉ

Ion concentration ProLab 4000100ba75646e02 01/2010Criteria The criteria of the stability control affect the reproducibility of the mea-sured values.

Strona 4

ProLab 4000 Ion concentration101ba75646e02 01/20109.2.3 Measurement settingsAll settings for ion selective measurements are done in the Measuring menu

Strona 5 - ProLab 4000 - Contents

Ion concentration ProLab 4000102ba75646e02 01/2010– Alternative TP () () : alternative temperature sen-sor is used( ) : no alternative temperatur

Strona 6

ProLab 4000 Ion concentration103ba75646e02 01/20109.3 CalibrationWhy calibrate? Ion-selective electrodes age and are temperature-dependent. This chang

Strona 7 - ProLab 4000 Contents

Ion concentration ProLab 4000104ba75646e02 01/2010ISE Cal This is the conventional two-point to nine-point calibration proce-dure that uses 2 to 9 fre

Strona 8

ProLab 4000 Ion concentration105ba75646e02 01/2010Calibration evaluation After calibrating, the meter automatically evaluates the calibration.Display

Strona 9

Ion concentration ProLab 4000106ba75646e02 01/20109.3.1 Carrying out calibrationUse two to nine standard solutions in ascending or descending order.Pr

Strona 10 - Contents ProLab 4000

ProLab 4000 Ion concentration107ba75646e02 01/20103 When measuring without temperature sensor:z Measure the temperature of the standard solution using

Strona 11 - 1Overview

Ion concentration ProLab 4000108ba75646e02 01/2010Continuing with two-point calibration9 Wait for the measurement with stability control to end.orTake

Strona 12 - 1.2 Keypad

ProLab 4000 Ion concentration109ba75646e02 01/2010Continuing with three-point to nine-point calibrationRepeat the steps 9 to 16 in the same way with t

Strona 13 - ProLab 4000 Overview

ProLab 4000 Overview11ba75646e02 01/20101Overview1.1 General featuresThe ProLab 4000 precision meter enables you to perform pH, ORP, conductivity and

Strona 14 - 1.3 Display

Ion concentration ProLab 4000110ba75646e02 01/20109.3.2 Calibration dataIn the Calibration menu, you can view the calibration data of ISE cali-bration

Strona 15 - 1.4 Socket field

ProLab 4000 Ion concentration111ba75646e02 01/20109.4 Blank value correctionIf the ion concentration in the sample is so small that it is not in the l

Strona 16 - Overview ProLab 4000

Ion concentration ProLab 4000112ba75646e02 01/20109.5 Reference measurementA reference measurement enables you to assign any concentration value to th

Strona 17

ProLab 4000 Ion concentration113ba75646e02 01/20109.6 Measuring with increment procedure (methods)9.6.1 Selecting the measuring methodThe following me

Strona 18

Ion concentration ProLab 4000114ba75646e02 01/20109.6.2 Standard additionIn the Standard addition procedure, a known amount of standard solu-tion is a

Strona 19 - 1.6 Electronic access control

ProLab 4000 Ion concentration115ba75646e02 01/20106 In the Sample register, enter the values for the sample volume (Sample volume) and the volume of t

Strona 20

Ion concentration ProLab 4000116ba75646e02 01/20109.6.3 Double standard additionIn the Double standard addition procedure, a known amount of stan-dard

Strona 21 - 2 Safety

ProLab 4000 Ion concentration117ba75646e02 01/20103 In the Sample register, enter the values for the sample volume (Sample volume) and the volume of t

Strona 22 - Safety ProLab 4000

Ion concentration ProLab 4000118ba75646e02 01/201010 Wait for the measurement with stability control to end.orTake over the measured value with [OK].

Strona 23 - 3 Commissioning

ProLab 4000 Ion concentration119ba75646e02 01/20109.6.4 Standard subtractionIn the "Standard Subtraction" procedure, a known amount of stand

Strona 24 - 3.2 Initial commissioning

Overview ProLab 400012ba75646e02 01/20101.2 KeypadKeys with dedicated functionsKeys with dedicated functions are labeled on the meter. In this operat

Strona 25 - ProLab 4000 Commissioning

Ion concentration ProLab 4000120ba75646e02 01/20106 In the Sample register, enter the values for the sample volume (Sample volume) and the volume of t

Strona 26 - 3.6 Connecting sensors

ProLab 4000 Ion concentration121ba75646e02 01/201016 Change the displayed unit of the measured value with [Unit] if necessary.17 If required, display

Strona 27

Ion concentration ProLab 4000122ba75646e02 01/20109.6.5 Sample additionIn the Sample addition procedure, a known amount of test sample is added to the

Strona 28 - Commissioning ProLab 4000

ProLab 4000 Ion concentration123ba75646e02 01/20106 In the Standard register, enter the values for Standard volume-Standard concentration and ISA/TISA

Strona 29

Ion concentration ProLab 4000124ba75646e02 01/201015 Change the displayed unit of the measured value with [Unit] if necessary.16 If required, display

Strona 30

ProLab 4000 Ion concentration125ba75646e02 01/20109.6.6 Sample subtractionIn the Sample subtraction procedure, a known amount of test sample is added

Strona 31

Ion concentration ProLab 4000126ba75646e02 01/20106 In the Standard register, enter the values for Standard volume-Standard concentration and ISA/TISA

Strona 32

ProLab 4000 Ion concentration127ba75646e02 01/201016 Change the displayed unit of the measured value with [Unit] if necessary.17 If required, display

Strona 33 - 4 Operating principles

Ion concentration ProLab 4000128ba75646e02 01/20109.6.7 Blank value addition (Standard addition with blank value correction)In the "Standard addi

Strona 34

ProLab 4000 Ion concentration129ba75646e02 01/2010Measuring6 In the Sample register, enter the following values:z the volume of the sample (Sample vol

Strona 35

ProLab 4000 Overview13ba75646e02 01/2010Alphanumeric keypad The keys of the alphanumeric keypad are used to enter numerals (<0...9>), letters (&

Strona 36

Ion concentration ProLab 4000130ba75646e02 01/201011 In the Standard register, enter the values for Standard volume and Standard concentration if nece

Strona 37 - 4.2 File system

ProLab 4000 Conductivity131ba75646d02 01/201010 Conductivity10.1 General informationYou can measure the following variables:z Conductivityz Specific r

Strona 38

Conductivity ProLab 4000132ba75646d02 01/201010.2.2 MeasuringYou can carry out conductivity measurements as follows:Selecting the displayed measured v

Strona 39

ProLab 4000 Conductivity133ba75646d02 01/2010Stability control The Stability control function checks the stability of the measurement signal. The stab

Strona 40 - 4.4 Navigation

Conductivity ProLab 4000134ba75646d02 01/201010.2.3 Measurement settingsThe following measurement settings are possible for conductivity mea-suring ce

Strona 41

ProLab 4000 Conductivity135ba75646d02 01/2010– Reference temperature – Tref 25– Tref 20Reference temperature– TDS factor setting... 0.40 ... 1.00 TDS

Strona 42

Conductivity ProLab 4000136ba75646d02 01/201010.2.4 Determining/setting the temperature compensation The conductivity of aqueous solutions is main

Strona 43

ProLab 4000 Conductivity137ba75646d02 01/2010Temperature compensa-tion for diluted solutionsFor the measurement of aqueous solutions z in a low concen

Strona 44

Conductivity ProLab 4000138ba75646d02 01/2010Temperature compensa-tion for concentrated solutionsFor the measurement of aqueous solutions z in a wide

Strona 45

ProLab 4000 Conductivity139ba75646d02 01/201010.2.5 Selecting the nonlinear temperature compensation TC nLF10.2.6 Selecting the linear temperature com

Strona 46

Overview ProLab 400014ba75646e02 01/20101.3 DisplayThe graphic color display can display three measured parameters and the current temperature value f

Strona 47 - 5 Access to the meter

Conductivity ProLab 4000140ba75646d02 01/201010.2.7 Selecting and adjusting the nonlinear temperature compensation TC nLin1 The nonlinear temperat

Strona 48

ProLab 4000 Conductivity141ba75646d02 01/20104 Select a test sample with permanently stored coefficients (NaCl, KCl, NaOH, HCl) or the freely adjustab

Strona 49 - 5.3 Password for login

Conductivity ProLab 4000142ba75646d02 01/201010.2.8 Selecting and determining the nonlinear temperature compensation TC nLin2 1 Highlight the chan

Strona 50 - 5.4 Lock

ProLab 4000 Conductivity143ba75646d02 01/2010NoteThe measuring method can only be started if Ref. temp., Start temp. and Temp. interval are selected s

Strona 51

Conductivity ProLab 4000144ba75646d02 01/201010.2.9 Selecting and determining the nonlinear temperature compensation TC nLin3 NoteThe measuring me

Strona 52

ProLab 4000 Conductivity145ba75646d02 01/20107 Warm or cool the measuring solution (standard 1) to the start temperature Start temp..As soon as the te

Strona 53

Conductivity ProLab 4000146ba75646d02 01/201010.2.10 Selecting and determining the nonlinear temperature compensation TC nLin4 Preparatory activit

Strona 54

ProLab 4000 Conductivity147ba75646d02 01/2010NoteThe measuring method can only be started if Ref. temp., Start temp. and Temp. interval are selected s

Strona 55 - 5.6 Lost your electronic key?

Conductivity ProLab 4000148ba75646d02 01/201010.2.11Switching off the temperature compensation (TC off)7 Slowly warm or cool the test sample (standard

Strona 56 - 6 System functions

ProLab 4000 Conductivity149ba75646d02 01/201010.3 Determining the cell constant (calibration in control standard)Why determine the cell constant?Due t

Strona 57 - ProLab 4000 System functions

ProLab 4000 Overview15ba75646e02 01/20101.4 Socket fieldConnections:1 Conductivity measuring cell (Cond Probe)2 pH/ISE/ORP electrode (pH/U/ISE Probe 2

Strona 58 - 6.3 Selecting the country

Conductivity ProLab 4000150ba75646d02 01/2010Sample record:Calibration evaluation After the calibration, the meter automatically evaluates the current

Strona 59 - 6.4 Setting the date and time

ProLab 4000 Conductivity151ba75646d02 01/201010.3.1 Determining the cell constant (calibration)Proceed as follows to determine the cell constant:1 Con

Strona 60 - System functions ProLab 4000

Conductivity ProLab 4000152ba75646d02 01/201010.3.2 Calibration settings and calibration data In the Calibration menu you can set the calibration inte

Strona 61 - 6.7 Temperature

ProLab 4000 Memory153ba75646e02 01/201011 MemoryWith the ProLab 4000 software you can store and recall all important measurement and configuration dat

Strona 62

Memory ProLab 4000154ba75646e02 01/201011.1.1 Creating and selecting a file for manually stored measurement datasetsCreating a file for manually store

Strona 63

ProLab 4000 Memory155ba75646e02 01/201011.1.2 Storing measurement datasets manuallyMeasurement datasets are stored in a file selected before (see page

Strona 64 - 6.9 Color assignment

Memory ProLab 4000156ba75646e02 01/201011.1.3 Automatically storing measurement datasetsThe file in which measurement datasets should be automatically

Strona 65 - 6.10 Reset

ProLab 4000 Memory157ba75646e02 01/2010 NoteThe progress of the automatic memory is indicated by a progress bar. The Automatic memory window can b

Strona 66

Memory ProLab 4000158ba75646e02 01/201011.1.4 Displaying and filtering stored measurement datasets1 Open the currently selected manual memory with <

Strona 67 - 6.11 List of open windows

ProLab 4000 Memory159ba75646e02 01/201011.1.5 Erasing measurement datasets4 Activate the selected filter criteria with [OK].The [Filter] option field

Strona 68

Overview ProLab 400016ba75646e02 01/20101.5 Automatic sensor recognitionThe automatic sensor recognition function enables z to operate a sensor with d

Strona 69 - 7.1 General information

Memory ProLab 4000160ba75646e02 01/201011.2 Calibration dataThe current calibration record is automatically stored after each valid calibration proced

Strona 70 - 7.2 Measuring the pH value

ProLab 4000 Memory161ba75646e02 01/201011.2.2 Managing the calibration dataThe calibration data is stored in the current calibration record (menu, Cal

Strona 71 - ProLab 4000 pH

Memory ProLab 4000162ba75646e02 01/201011.3 Recorder dataInformation on the recording, storing, displaying and deleting of mea-sured value curves is g

Strona 72

ProLab 4000 Recorder163ba75646e02 01/201012 RecorderThe recorder represents measured values graphically in a system of coordinates. The recorder can r

Strona 73

Recorder ProLab 4000164ba75646e02 01/201012.1 Recording the measured value curveThe recorder records the course of the measured values curve and the c

Strona 74 - 7.3 Calibration

ProLab 4000 Recorder165ba75646e02 01/201012.2 Recording the curve of the measured temperature value If the recorder was started without recor

Strona 75

Recorder ProLab 4000166ba75646e02 01/201012.3 Storing recorder dataThe recorder records the course of the measured values curve and the course of the

Strona 76

ProLab 4000 Recorder167ba75646e02 01/201012.4 Displaying stored recorder dataOpening the offline recorderDisplaying a different recorder file1 Open th

Strona 77

Recorder ProLab 4000168ba75646e02 01/201012.5 Settings for the recorderIn the Recorder settings dialog box, the appearance of your recorder is defined

Strona 78 - HAT TO DO IF... (page 189)

ProLab 4000 Recorder169ba75646e02 01/201012.5.1 Defining the recording intervalThe Recording interval defines at which intervals the measured values a

Strona 79

ProLab 4000 Overview17ba75646e02 01/20101.5.1 ID sensorsSI Analytics GmbH ID sensors support the automatic sensor recogni-tion function. Their sensor

Strona 80

Recorder ProLab 4000170ba75646e02 01/20101 Open the recorder menu, Settings...Settings...The Recorder settings window pops up.2 Select the Measured v

Strona 81

ProLab 4000 Recorder171ba75646e02 01/201012.5.3 Defining and activating limit valuesWith the limit values you define the measured values at which the

Strona 82

Recorder ProLab 4000172ba75646e02 01/201012.5.4 Defining the colors for the recorder representation The appearance of the recorder is determined w

Strona 83

ProLab 4000 Recorder173ba75646e02 01/201012.6 Changing the displayed axis interceptThe standard settings for the displayed axis intercepts are defined

Strona 84

Recorder ProLab 4000174ba75646e02 01/2010Zooming an axis interceptThe axis intercepts of the recorder can be zoomed as follows (time axis and measured

Strona 85

ProLab 4000 Recorder175ba75646e02 01/201012.7 Displaying exact measurement data at a point (cursor)In the recorder window, you can display the exact m

Strona 86 - Note

Recorder ProLab 4000176ba75646e02 01/2010

Strona 87

ProLab 4000 Transmitting data177ba75646e02 01/201013 Transmitting data13.1 Printing current measurement dataTo which interface the measurement data sh

Strona 88

Transmitting data ProLab 4000178ba75646e02 01/201013.2.2 USB-B interface (USB Device)Connect the interface to the PC via the supplied Z875 USB cable.

Strona 89 - 8 ORP voltage

ProLab 4000 Transmitting data179ba75646e02 01/201013.2.3 USB-A interface (USB Host)Connect the USB-A interface (USB Host) of the ProLab 4000 to the pr

Strona 90 - 8.2 Measuring the ORP

Overview ProLab 400018ba75646e02 01/20101.5.2 Sensor data from ID sensorsID sensors transmit the following sensor data to the meter:z SENSOR ID– Senso

Strona 91 - ProLab 4000 ORP voltage

Transmitting data ProLab 4000180ba75646e02 01/201013.3 Data backup13.3.1 Data backup by the administratorTo back up important data against being lost

Strona 92 - ORP voltage ProLab 4000

ProLab 4000 Transmitting data181ba75646e02 01/201013.3.2 Viewing backed up dataThe external USB medium is displayed within the directory of each user.

Strona 93

Configurations ProLab 4000182ba75646e02 01/201014 ConfigurationsAs the administrator, you can create and assign a configuration to each user (see page

Strona 94

ProLab 4000 Configurations183ba75646e02 01/201014.1 Viewing a configurationA configuration viewer is available so the current configuration settings c

Strona 95

Configurations ProLab 4000184ba75646e02 01/201014.2 Creating a new configurationWhen creating a configuration, the current settings of the meter for e

Strona 96

ProLab 4000 Configurations185ba75646e02 01/201014.3 Loading a configurationAll users who can create a configuration can also load it.When loading a co

Strona 97 - 9 Ion concentration

Configurations ProLab 4000186ba75646e02 01/201014.4 Deleting a configurationAll users who can create a configuration can also delete it in their direc

Strona 98

ProLab 4000 Maintenance, cleaning, disposal187ba75646e02 01/201015 Maintenance, cleaning, disposal15.1 MaintenanceThe only maintenance activity requir

Strona 99

Maintenance, cleaning, disposal ProLab 4000188ba75646e02 01/201015.2 CleaningOccasionally wipe the outside of the meter with a damp, lint-free cloth.

Strona 100 - Ion concentration ProLab 4000

ProLab 4000 What to do if...189ba75646e02 01/201016 What to do if...16.1 pH and ORP measurementError message Err1Error message Err2Error message Err4E

Strona 101 - ProLab 4000 Ion concentration

ProLab 4000 Overview19ba75646e02 01/20101.6 Electronic access controlThe ProLab 4000 always documents measurement and calibration data together with a

Strona 102

What to do if... ProLab 4000190ba75646e02 01/2010No stable measured value Obviously incorrect measured values– Electrode broken – Replace the electr

Strona 103 - 9.3 Calibration

ProLab 4000 What to do if...191ba75646e02 01/2010CalClock on red background16.2 Ion selective measurementError message Err1 Obviously incorrect measur

Strona 104

What to do if... ProLab 4000192ba75646e02 01/2010Warning [TempErr] Warning [ISEErr] 16.3 Conductivity measurementConductivity measuring cell was not

Strona 105

ProLab 4000 What to do if...193ba75646e02 01/201016.4 General errorsMeter does not react to keystrokeTime is lostCause Remedy– Operating condition und

Strona 106

What to do if... ProLab 4000194ba75646e02 01/2010

Strona 107

ProLab 4000 Technical Data195ba75646e02 01/201017 Technical Data17.1 General dataDimensions approx. 240 x 280 x 70 mmWeight approx. 2.5 kg (without po

Strona 108

Technical Data ProLab 4000196ba75646e02 01/2010RS232 serial interface Connection of the PC cable (Z390) or printer cable (Z393)Baud rate adjustable:

Strona 109

ProLab 4000 Technical Data197ba75646e02 01/2010Instrument safety EC guideline 2006/95/ECEN 61010-1 class AANSI/UL 61010-1CAN/CSA-C22.2 No. 61010-1Rad

Strona 110

Technical Data ProLab 4000198ba75646e02 01/201017.2 Measuring ranges, resolution, accuracy17.2.1 Temperature17.2.2 pH/ORPMeasuring ranges, resolutionV

Strona 111 - 9.4 Blank value correction

ProLab 4000 Technical Data199ba75646e02 01/201017.2.3 ISE17.2.4 ConductivityMeasuring ranges, resolutionVariable Measuring range ResolutionISE [mg/l,

Strona 112 - 9.5 Reference measurement

Accuracy when going to pressThe use of advanced technology and the high quality standard of our instruments are the result of a continuous development

Strona 113

Overview ProLab 400020ba75646e02 01/2010

Strona 114

Technical Data ProLab 4000200ba75646e02 01/2010Cell constants Cell constant C ValuesCan be calibrated in the ranges0.450 ... 0.500 cm-10.585 ... 0.71

Strona 115

ProLab 4000 Lists201ba75646e02 01/201018 ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviation

Strona 116

Lists ProLab 4000202ba75646e02 01/2010AbbreviationsConductivity value (international γ)°C Temperature unit, degrees Celsius°F Temperature unit, degree

Strona 117

ProLab 4000 Lists203ba75646e02 01/2010Status display indicators[SC] Measured value (Hold) or Measurement with stability control[ARng] Automatic range

Strona 118

Lists ProLab 4000204ba75646e02 01/2010GlossaryAdjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differ

Strona 119

ProLab 4000 Lists205ba75646e02 01/2010Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g solvent.Offset potential The measu

Strona 120

Lists ProLab 4000206ba75646e02 01/2010Temperature compensation Name of a function that considers the temperature influence on the measurement and conv

Strona 121

ProLab 4000 Lists207ba75646e02 01/2010Trademarks usedTrademark Owner of the trademarkMerck Merck KGaASCHOTT Schott Glas AGMettler Toledo Mettler Toled

Strona 122

Lists ProLab 4000208ba75646e02 01/2010

Strona 123

ProLab 4000 Lists 209Indexba75646e02 01/2010AAuthorized use ...21BBattery compartment ...

Strona 124

ProLab 4000 Safety21ba75646e02 01/20102 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operati

Strona 125

Index ProLab 4000210ba75646e02 01/2010Reset ... 65RS232 interface ... 178,

Strona 126

ProLab 4000 Appendix211ba75646e02 01/2010AppendixA.19 Firmware updateGeneral information With the "Update_ProLab3000_4000" program and a PC

Strona 127

Appendix ProLab 4000212ba75646e02 01/2010Program installationProgram start Firmware updateAfter switching the meter off and on you can check whether t

Strona 128

ProLab 4000 Appendix213ba75646e02 01/2010A.20 MenusA.20.1 Main menuFile MemoryFile – New configuration...– Select configuration...– View configuration

Strona 129

Appendix ProLab 4000214ba75646e02 01/2010SystemWindowUser Help System– Data output...– Baud– Output format (ASCII, CSV)–– Header– Autom. printout...–

Strona 130

ProLab 4000 Appendix215ba75646e02 01/2010A.20.2 pH sensor menuMeasuringCalibrationMeasuring–pH resolution–High– Medium– Low– Alternative TP– Setting o

Strona 131 - 10 Conductivity

Appendix ProLab 4000216ba75646e02 01/2010A.20.3 U sensor menuMeasuringA.20.4 dU sensor menuMeasuringMeasuring– High resolution– Alternative TP– Settin

Strona 132 - Conductivity ProLab 4000

ProLab 4000 Appendix217ba75646e02 01/2010A.20.5 ISE sensor menuMeasuringCalibrationMeasuring– ISE settings–Units– Ion type– Blank value correction– Re

Strona 133 - ProLab 4000 Conductivity

Appendix ProLab 4000218ba75646e02 01/2010A.20.6 Cond sensor menuMeasuringCalibrationMeasuring– Cell constant– Calibrated cell constant– Adjustable cel

Strona 134

SI Analytics GmbHPostfach 2443D-55014 MainzHattenbergstr. 10D-55122 MainzEmail: support@Telefon +49 (0) 61 31/66 5111Telefax +49 (0) 61 31/66 5001si-a

Strona 135

Safety ProLab 400022ba75646e02 01/20102.2 General safety instructionsThis instrument is built and inspected according to the relevant guide-lines and

Strona 136

ProLab 4000 Commissioning23ba75646e02 01/20103 Commissioning3.1 Scope of deliveryz ProLab 4000 laboratory meterz Power packz 4 batteries 1.5 V Micro t

Strona 137

Commissioning ProLab 400024ba75646e02 01/20103.2 Initial commissioningPerform the following activities:z Insert the batteries (see page 187)z Connect

Strona 138

ProLab 4000 Commissioning25ba75646e02 01/20103.4 Switching on the ProLab 4000The ProLab 4000 indicates the Standby display.1 Switch the meter on with

Strona 139

Commissioning ProLab 400026ba75646e02 01/20103.5 Connecting the USB mouseYou can connect the USB mouse included in the scope of delivery to the USB-A

Strona 140

ProLab 4000 Commissioning27ba75646e02 01/20103.6.1 Connecting an ID sensorID sensor with temperature sensorID sensor without temperature sensor1 Conne

Strona 141

Commissioning ProLab 400028ba75646e02 01/20103.6.2 Connecting a non ID sensor3.7 Connecting optional accessoriesThe ProLab 4000 has several interfaces

Strona 142

ProLab 4000 Commissioning29ba75646e02 01/20103.7.1 RS232 interface (serial port)PC and printer can be connected via the RS232 interface.Data that is t

Strona 143

ProLab 40003ba75646e02 01/2010KONFORMITÄTSERKLÄRUNGDECLARATION OF CONFORMITYDÉCLARATION DE CONFORMITÉWir erklären in alleiniger Verantwortung, dass da

Strona 144

Commissioning ProLab 400030ba75646e02 01/20103.7.2 USB-B interface (USB Device)You can connect a PC to the meter via the USB-B interface(USB Device).

Strona 145

ProLab 4000 Commissioning31ba75646e02 01/20103.7.3 USB-A interface (USB Host)You can connect the following devices to the USB-A interface (USB Host) a

Strona 146

Commissioning ProLab 400032ba75646e02 01/2010

Strona 147

ProLab 4000 Operating principles33ba75646e02 01/20104 Operating principlesThis section contains basic information on the operation of the ProLab 4000.

Strona 148

Operating principles ProLab 400034ba75646e02 01/2010Display Depending on the operating situation, the color display shows mea-surement data, calibrati

Strona 149

ProLab 4000 Operating principles35ba75646e02 01/2010In the main menu, there are sensor independent functions and settings: z Manage configurations (Fi

Strona 150

Operating principles ProLab 400036ba75646e02 01/2010Channel (2) In the measured value display, for each connected sensor you can see a channel with sp

Strona 151

ProLab 4000 Operating principles37ba75646e02 01/20104.2 File systemFiles With the ProLab 4000 you can store data in files just like using a PC. Differ

Strona 152

Operating principles ProLab 400038ba75646e02 01/20104.3 Entry of numerals, letters and charactersNumerals, letters, punctuation marks and special char

Strona 153 - 11 Memory

ProLab 4000 Operating principles39ba75646e02 01/2010Mouse operation:With the mouse, you have in all entry fields the additional option of showing a ke

Strona 154 - Memory ProLab 4000

ProLab 40004ba75646e02 01/2010

Strona 155 - ProLab 4000 Memory

Operating principles ProLab 400040ba75646e02 01/20104.4 NavigationThe dialog with the meter takes place through menus and dialog boxes as usual with a

Strona 156

ProLab 4000 Operating principles41ba75646e02 01/20104.4.1 Navigation in the measured value displayThe display shows, for example, the measurement data

Strona 157

Operating principles ProLab 400042ba75646e02 01/20104.4.2 Navigation in menusA menu item contains submenus, executes a function or opens a dialog box.

Strona 158

ProLab 4000 Operating principles43ba75646e02 01/2010Mouse operation:5 Highlight a menu item with <W><X> and <S><T>.The highlig

Strona 159

Operating principles ProLab 400044ba75646e02 01/20104.4.3 Navigation in dialog boxesDialog boxes contain further subelements such as register cards, s

Strona 160 - 11.2 Calibration data

ProLab 4000 Operating principles45ba75646e02 01/20104.4.4 Navigation in the file selection dialog boxThe file selection dialog is opened if a file is

Strona 161

Operating principles ProLab 400046ba75646e02 01/2010Operating example: Navigation and settings in file selection dialogsMouse operation:1 Highlight an

Strona 162 - 11.4 Configuration data

ProLab 4000 Access to the meter47ba75646e02 01/20105 Access to the meter5.1 Switch the meter on and offSwitching onSwitching off5.2 Login as a userAft

Strona 163 - 12 Recorder

Access to the meter ProLab 400048ba75646e02 01/2010Login with electronic keyThe Login dialog box displays the instruction: Place the electronic key o

Strona 164 - Recorder ProLab 4000

ProLab 4000 Access to the meter49ba75646e02 01/20105.3 Password for loginIf the used of a password for login to the meter was set by the admin-istrato

Strona 165 - ProLab 4000 Recorder

ProLab 4000 Contents5ba75646e02 01/2010ProLab 4000 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111.

Strona 166 - 12.3 Storing recorder data

Access to the meter ProLab 400050ba75646e02 01/20105.4 LockDuring operation (e.g. if the automatic storing function is active), the activated lock pre

Strona 167

ProLab 4000 Access to the meter51ba75646e02 01/20105.5 Access control and user rightsYou as the administrator define the type of access (assess contro

Strona 168

Access to the meter ProLab 400052ba75646e02 01/20105.5.1 Access controlThe administrator defines the type of access to the meter. The setting applies

Strona 169

ProLab 4000 Access to the meter53ba75646e02 01/20105.5.2 User management and assigning user rightsThe ProLab 4000 has basic user management functions.

Strona 170

Access to the meter ProLab 400054ba75646e02 01/2010Adding a new userAssigning a configuration to a userErasing a registered user1 Enter a new user nam

Strona 171

ProLab 4000 Access to the meter55ba75646e02 01/2010Erasing the password for a registered userHiding / displaying a userExiting the user management5.6

Strona 172

System functions ProLab 400056ba75646e02 01/20106 System functionsSystem functions are all functions that are independent of the probe. 6.1 Overview:

Strona 173

ProLab 4000 System functions57ba75646e02 01/2010– Signal On () Switches on or off the acoustic signal. The signal sounds e.g. when a stable measured

Strona 174

System functions ProLab 400058ba75646e02 01/20106.2 Selecting the languageMenus can be displayed in different languages. The language is selected in t

Strona 175

ProLab 4000 System functions59ba75646e02 01/20106.4 Setting the date and timeThe data and time are displayed in the local format, according to the Cou

Strona 176

Contents ProLab 40006ba75646e02 01/20105 Access to the meter . . . . . . . . . . . . . . . . . . . . . . . . . . . 475.1 Switch the meter on and off

Strona 177 - 13 Transmitting data

System functions ProLab 400060ba75646e02 01/20106.5 Selecting the channels for measured value displayIn the Window / Channels... you can define which

Strona 178 - Transmitting data ProLab 4000

ProLab 4000 System functions61ba75646e02 01/20106.7 TemperatureThe temperature affects the measurement results. You can select the unit of the tempera

Strona 179 - ProLab 4000 Transmitting data

System functions ProLab 400062ba75646e02 01/20106.7.2 Temperature measurementFor reproducible measurements it is essential to determine the corre-spon

Strona 180 - 13.3 Data backup

ProLab 4000 System functions63ba75646e02 01/20106.8 Automatic stability controlThe Autom. stability control function continuously checks the stability

Strona 181

System functions ProLab 400064ba75646e02 01/20106.9 Color assignmentFor easy differentiation, the following conditions are highlighted by a background

Strona 182 - 14 Configurations

ProLab 4000 System functions65ba75646e02 01/20106.10 ResetYou can reset (initialize) all sensor settings and sensor-independent settings separately fr

Strona 183 - 14.1 Viewing a configuration

System functions ProLab 400066ba75646e02 01/20106.10.2 Resetting the system settingsThe following system settings are reset to the default settings wi

Strona 184 - Configurations ProLab 4000

ProLab 4000 System functions67ba75646e02 01/20106.11 List of open windowsThe Window menu item in the main menu provides an overview of the open window

Strona 185 - 14.3 Loading a configuration

System functions ProLab 400068ba75646e02 01/2010

Strona 186 - 14.4 Deleting a configuration

ProLab 4000 pH69ba75646e02 01/20107pH7.1 General informationTemperature measurementFor reproducible pH measurements, it is essential to measure the te

Strona 187 - 15.1 Maintenance

ProLab 4000 Contents7ba75646e02 01/20108 ORP voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 898.1 General information. .

Strona 188 - 15.3 Disposal

pH ProLab 400070ba75646e02 01/20107.2 Measuring the pH value7.2.1 Preparatory activitiesPerform the following preparatory activities when you want to

Strona 189 - 16 What to do if

ProLab 4000 pH71ba75646e02 01/20107.2.2 MeasuringStability control The Stability control function checks the stability of the measurement signal. The

Strona 190 - What to do if... ProLab 4000

pH ProLab 400072ba75646e02 01/2010Criteria With identical measurement conditions, the following applies: NoteYou can terminate prematurely the Sta

Strona 191 - ± 2 °C temperature differ

ProLab 4000 pH73ba75646e02 01/20107.2.3 Measurement settingsAll settings for pH measurements are in the Measuring menu.Menu item Setting DescriptionMe

Strona 192 - 16.3 Conductivity measurement

pH ProLab 400074ba75646e02 01/20107.3 CalibrationWhy calibrate? pH electrodes age. This changes the zero point (asymmetry) and slope of the pH electro

Strona 193 - 16.4 General errors

ProLab 4000 pH75ba75646e02 01/20108 DIN 19267 * 1,0904,6506,7909,23025 °C9 Mettler EU 1,6794,0037,00210,01325 °C10 Mettler US * 1,9954,0057,0029,20825

Strona 194

pH ProLab 400076ba75646e02 01/2010* Brand names or trade names are trademarks of their respective owners protected by law (see page 207).Calibration p

Strona 195 - 17 Technical Data

ProLab 4000 pH77ba75646e02 01/2010Stability control The calibration procedure automatically activates the stability control function. The current meas

Strona 196 - Technical Data ProLab 4000

pH ProLab 400078ba75646e02 01/2010Calibration evaluation After calibrating, the meter automatically evaluates the calibration. The zero point and slop

Strona 197 - ProLab 4000 Technical Data

ProLab 4000 pH79ba75646e02 01/20107.3.1 Carrying out a calibration procedure (example: AutoCal DIN) Make sure the Calibration / pH calibratio

Strona 198

Contents ProLab 40008ba75646e02 01/201010.2.10 Selecting and determining the nonlinear temperature compensation TC nLin4 . . . . . 14610.2.11 Switchi

Strona 199

pH ProLab 400080ba75646e02 01/20101 Recognized nominal buffer value (referring to 25 °C)2 Current electrode voltage3 Current temperature value4 Select

Strona 200

ProLab 4000 pH81ba75646e02 01/2010Continuing with two-point calibration (Buffer: DIN)9 Using [OK], confirm the Finish with 1-point calibration infotex

Strona 201 - 18 Lists

pH ProLab 400082ba75646e02 01/2010Continuing with three-point to five-point calibration (Buffer ) DIN) NoteA calibration line is determined for the

Strona 202 - Lists ProLab 4000

ProLab 4000 pH83ba75646e02 01/20107.3.2 Carrying out a calibration procedure (example: VariCal)Make sure the Calibration / pH calibration type / VariC

Strona 203 - ProLab 4000 Lists

pH ProLab 400084ba75646e02 01/20101 Recognized nominal buffer value (referring to 25 °C)2 Current electrode voltage3 Current temperature value4 Select

Strona 204

ProLab 4000 pH85ba75646e02 01/2010Continuing with two-point calibration8 Using [OK], confirm the Finish with 1-point calibration infotext and take ove

Strona 205

pH ProLab 400086ba75646e02 01/2010 NoteA calibration line is determined for the two-point calibration.

Strona 206

ProLab 4000 pH87ba75646e02 01/20107.3.3 Calibration settings and calibration dataIn the Calibration menu, you can z set the calibration interval, z de

Strona 207

pH ProLab 400088ba75646e02 01/20107.3.4 Calibration intervalThe calibration interval and calibration evaluation are indicated on the display as the Ca

Strona 208

ProLab 4000 ORP voltage89ba75646e02 01/20108 ORP voltage8.1 General informationYou can measure the following variables:z ORP [mV]z Relative ORP [mV]Te

Strona 209

ProLab 4000 Contents9ba75646e02 01/201013.3 Data backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18013.3.1 Data backup by t

Strona 210 - Index ProLab 4000

ORP voltage ProLab 400090ba75646e02 01/20108.2 Measuring the ORP8.2.1 Preparatory activitiesPerform the following preparatory activities when you want

Strona 211 - Appendix

ProLab 4000 ORP voltage91ba75646e02 01/20108.2.2 MeasuringStability control TheStability control function checks the stability of the measurement sign

Strona 212 - Appendix ProLab 4000

ORP voltage ProLab 400092ba75646e02 01/2010Criteria With identical measurement conditions, the following applies:4With <OK> activate the Stabili

Strona 213 - A.20 Menus

ProLab 4000 ORP voltage93ba75646e02 01/20108.2.3 Measuring the relative ORPTo measure the difference of the ORPs of two solutions it is best to use tw

Strona 214

ORP voltage ProLab 400094ba75646e02 01/2010As an alternative you can also determine the difference of the ORPs of two solutions with one ORP electrode

Strona 215 - ProLab 4000 Appendix

ProLab 4000 ORP voltage95ba75646e02 01/2010Criteria With identical measurement conditions, the following applies:1 In the measured value display, mark

Strona 216

ORP voltage ProLab 400096ba75646e02 01/20108.2.4 Measurement settingsThe following settings are possible for ORP measurements in the Mea-suring menu:M

Strona 217

ProLab 4000 Ion concentration97ba75646e02 01/20109 Ion concentration9.1 General informationTemperature measurement with ion selective measurementsThe

Strona 218

Ion concentration ProLab 400098ba75646e02 01/20109.2 Measuring the ion concentration 9.2.1 Preparatory activitiesPerform the following preparatory act

Strona 219

ProLab 4000 Ion concentration99ba75646e02 01/20109.2.2 MeasuringTemperature for calibrating and measuringFor precise ion selective measurements, the t