Xylem Lab 860 Instrukcja Użytkownika Strona 1

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Narzędzia Xylem Lab 860. Xylem Lab 860 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Operating manual

ba75554e05 01/2010Lab 860 Operating manualLaboratory pH meter8422996pH°CTPLab 860

Strona 2 - © 2010, SI Analytics GmbH

Overview Lab 86010ba75554e05 01/20101.3 Socket fieldCAUTIONOnly connect sensors to the meter that cannot return any voltages or currents that are not

Strona 3 - DÉCLARATION DE CONFORMITÉ

Lab 860 Safety11ba75554e05 01/20102 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation a

Strona 4

Safety Lab 86012ba75554e05 01/20102.1 Authorized useThis meter is authorized exclusively for pH and ORP measurements in the laboratory. The technical

Strona 5 - Lab 860 - Contents

Lab 860 Safety13ba75554e05 01/2010Safe operation If safe operation is no longer possible, the instrument must be taken out of service and secured agai

Strona 6 - Contents Lab 860

Safety Lab 86014ba75554e05 01/2010

Strona 7 - 1Overview

Lab 860 Commissioning15ba75554e05 01/20103 Commissioning3.1 Scope of deliveryz Lab 860 laboratory meter z Power packz 4 batteries 1.5 V Mignon type AA

Strona 8 - 1.1 Keypad

Commissioning Lab 86016ba75554e05 01/2010CAUTIONMake sure that the poles of the batteries are the right way round.The ± signs on the batteries must co

Strona 9 - 1.2 Display

Lab 860 Operation17ba75554e05 01/20104 Operation4.1 Switching on the meterNoteThe meter has an energy saving feature to avoid unnecessary battery depl

Strona 10 - 1.3 Socket field

Operation Lab 86018ba75554e05 01/20104.2 General operating principlesThis section contains basic information on the operation of the Lab 860. 4.2.1 Op

Strona 11 - 2 Safety

Lab 860 Operation19ba75554e05 01/20104.3 MeasuringPreparatory activities Perform the following preparatory activities when you want to measure:NoteInc

Strona 12 - 2.1 Authorized use

Lab 8602ba75554e05 01/2010Accuracy when going topressThe use of advanced technology and the high quality standard of our instruments are the result of

Strona 13 - Lab 860 Safety

Operation Lab 86020ba75554e05 01/20104.3.1 Measuring the pH valueStability control SC(drift control)The stability control function (drift control) che

Strona 14 - Safety Lab 860

Lab 860 Operation21ba75554e05 01/20104.3.2 Measuring the ORPThe meter can, in conjunction with an ORP electrode, measure the ORP (mV) of a solution. N

Strona 15 - 3 Commissioning

Operation Lab 86022ba75554e05 01/20104.4 CalibrationWhy calibrate? pH electrodes age. This changes the zero point (asymmetry) and slope of the pH elec

Strona 16 - Commissioning Lab 860

Lab 860 Operation23ba75554e05 01/20108 Mettler Toledo TEC * 1.9954.0057.0029.20825 °C9 Fisher * 2.0074.0027.00410.00225 °C10 Fluka BS * 4.0066.9848.95

Strona 17 - 4 Operation

Operation Lab 86024ba75554e05 01/2010NoteThe buffer set is selected in the menu for measurement settings (see page 46).A list of the stored buffer set

Strona 18 - Operation Lab 860

Lab 860 Operation25ba75554e05 01/2010Displaying thecalibration dataYou can display the data of the last calibration (see page 36). Subsequently, you c

Strona 19 - 4.3 Measuring

Operation Lab 86026ba75554e05 01/2010Preparatory activitiesCalError CalError< -30 or > 30... -62 or ... -50 Eliminate the error according to pag

Strona 20

Lab 860 Operation27ba75554e05 01/20104.4.1 Calibration interval (Int.C)The calibration interval and calibration evaluation are indicated on the displa

Strona 21 - Lab 860 Operation

Operation Lab 86028ba75554e05 01/20104.4.2 Automatic calibration (AutoCal)Use any one to five buffer solutions of the selected buffer set.NoteThe step

Strona 22 - 4.4 Calibration

Lab 860 Operation29ba75554e05 01/2010NoteHere you can cancel the calibration procedure with <MODE>. This corresponds to a single-point calibrati

Strona 23

Lab 8603ba75554e05 01/2010KONFORMITÄTSERKLÄRUNGDECLARATION OF CONFORMITYDÉCLARATION DE CONFORMITÉWir erklären in alleiniger Verantwortung, dass das Pr

Strona 24

Operation Lab 86030ba75554e05 01/2010NoteHere you can cancel the calibration procedure with <MODE>. This corresponds to a two-point calibration.

Strona 25 - _> key

Lab 860 Operation31ba75554e05 01/20104.5 SavingThe pH meter has an internal data memory. It can store up to 800 datasets. A complete dataset consists

Strona 26

Operation Lab 86032ba75554e05 01/2010 Message StoFull This message appears when all of the 800 memory locations are occupied.When the memory is full,

Strona 27

Lab 860 Operation33ba75554e05 01/20104.5.2 Automatic storage at intervals (AutoStore)In order to store measured values automatically at certain time i

Strona 28

Operation Lab 86034ba75554e05 01/2010Switching off theautomatic storingfunctionYou can terminate the automatic storing function in the following ways:

Strona 29

Lab 860 Operation35ba75554e05 01/20104.5.3 Downloading the data memoryYou can download stored measurement datasets:z to the displayz to the interface

Strona 30 - <▲><▼>

Operation Lab 86036ba75554e05 01/20104.5.4 Erasing the data memoryYou can erase all stored measurement datasets.4.5.5 Downloading calibration dataYou

Strona 31 - 4.5 Saving

Lab 860 Operation37ba75554e05 01/2010Download to display viacalibration menuWhile the calibration data is being displayed, you can:z press <OK>

Strona 32

Operation Lab 86038ba75554e05 01/20104.6 Transmitting dataThe meter has two interfaces: z RS232 interface (serial port)z USB interface (device)Via bot

Strona 33

Lab 860 Operation39ba75554e05 01/20104.6.1 Options for data transmissionVia the USB interface you can transmit data to a PC. Via the RS 232 interface,

Strona 34

Lab 8604ba75554e05 01/2010

Strona 35

Operation Lab 86040ba75554e05 01/20104.6.2 Automatically downloading measurement datasets at intervalsIn order to automatically download to the interf

Strona 36

Lab 860 Operation41ba75554e05 01/2010Socket assignment(RS232)4.6.6 USB interface (device)Connect the interface to the PC via the supplied Z875 USB cab

Strona 37

Operation Lab 86042ba75554e05 01/2010operating manual.

Strona 38 - 4.6 Transmitting data

Lab 860 Operation43ba75554e05 01/20104.7 SettingsYou can adapt the meter to your individual requirements. The settings are done in the following menus

Strona 39 - _>. Then you

Operation Lab 86044ba75554e05 01/20104.7.1 System settingsThe default setting is printed in bold. Baud rate (Baud) Switch-off interval(t.OFF)Baud ra

Strona 40

Lab 860 Operation45ba75554e05 01/2010 Date and time4 Set the switch-off interval with <▲><▼>.5 Confirm with <OK>.Day.Month, the sett

Strona 41

Operation Lab 86046ba75554e05 01/20104.7.2 Measurement settingsThese settings apply to calibration and measurement (the default setting is printed in

Strona 42

Lab 860 Operation47ba75554e05 01/2010Display duringcalibration(bUFF)Unit of the value for theslope (SLO)4 Using <▲><▼>, select the display

Strona 43 - 4.7 Settings

Operation Lab 86048ba75554e05 01/2010Unit of the value for thezero point (ASY)Resolution(rES)8 Using <▲><▼>, select the unit for the zero

Strona 44

Lab 860 Operation49ba75554e05 01/2010Temperature unit (Unit)Calibrationinterval(Int.C) 12 Using <▲><▼>, toggle between °C and °F.13 Confir

Strona 45

Lab 860 Contents5ba75554e05 01/2010Lab 860 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71.1 Keypa

Strona 46

Operation Lab 86050ba75554e05 01/20104.7.3 Interval for automatic data transmissionThe interval for automatic data transmission serves to transmit the

Strona 47

Lab 860 Operation51ba75554e05 01/20104.8 Reset4.8.1 Resetting calibration valuesThis function resets the calibration values to the default condition.

Strona 48

Operation Lab 86052ba75554e05 01/20104.8.2 Resetting all meter settingsThis function resets to the default condition all meter settings including the

Strona 49 - toggle between °C and °F

Lab 860 Maintenance, cleaning, disposal53ba75554e05 01/20105 Maintenance, cleaning, disposal5.1 MaintenanceThe only maintenance activity required is r

Strona 50 - to select an interval

Maintenance, cleaning, disposal Lab 86054ba75554e05 01/20105.2 CleaningOccasionally wipe the outside of the meter with a damp, lint-free cloth. Disinf

Strona 51 - 4.8 Reset

Lab 860 What to do if...55ba75554e05 01/20106 What to do if...Error messageErr1Error messageErr2Error messageErr4Error messageCalErrorCause RemedypH e

Strona 52

What to do if... Lab 86056ba75554e05 01/2010No stable measuredvalue CalClock flashesDisplay,LoBat– Incorrect buffer solutions – Change calibration pro

Strona 53 - 5.1 Maintenance

Lab 860 What to do if...57ba75554e05 01/2010DisplaytoObviously incorrectmeasured valuesInstrument does notreact to keystrokeYou want to know whichsoft

Strona 54 - 5.4 Disposal

What to do if... Lab 86058ba75554e05 01/2010

Strona 55 - 6 What to do if

Lab 860 Technical data59ba75554e05 01/2010Stand October 30, 2009Translation of the legally binding German version7 Technical data7.1 General dataDimen

Strona 56 - What to do if... Lab 860

Contents Lab 8606ba75554e05 01/20104.7 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 434.7.1 System settings .

Strona 57 - Lab 860 What to do if

Technical data Lab 86060ba75554e05 01/2010Stand October 30, 2009Translation of the legally binding German versionGuidelinesand norms usedEMC EC guidel

Strona 58

Lab 860 Technical data61ba75554e05 01/2010Stand October 30, 2009Translation of the legally binding German version7.2 Measuring ranges, resolution, acc

Strona 59 - 7 Technical data

Technical data Lab 86062ba75554e05 01/2010Stand October 30, 2009Translation of the legally binding German version

Strona 60 - Technical data Lab 860

Lab 860 Lists63ba75554e05 01/20108ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviations expla

Strona 61 - Lab 860 Technical data

Lists Lab 86064ba75554e05 01/2010Abbreviations°C Temperature unit °Celsius°F Temperature unit, °FahrenheitASY Zero point (asymmetry)AutoCal Automatic

Strona 62

Lab 860 Lists65ba75554e05 01/2010GlossaryAsymmetry see zero pointResolution Smallest difference between two measured values that can be displayed by a

Strona 63 - Lab 860 Lists

Lists Lab 86066ba75554e05 01/2010Potentiometry Name of a measuring technique. The signal (depending on the measured parameter) of the electrode is the

Strona 64 - Lists Lab 860

Lab 860 Lists67ba75554e05 01/2010Trademarks usedTrademark Owner of the trademarkMerck Merck KGaAMettler Toledo Mettler ToledoFisher Fisher Scientific

Strona 65

Lists Lab 86068ba75554e05 01/2010

Strona 66

Lab 860 ListsIndex69ba75554e05 01/2010AAuthorized use . . . . . . . . . . . . . . . . . . . . . . 12AutoCal . . . . . . . . . . . . . . . . . . . .

Strona 67

Lab 860 Overview7ba75554e05 01/20101OverviewThe compact Lab 860 precision pH meter enables you to perform pH measurements rapidly and reliably. The La

Strona 68

Lists Lab 86070ba75554e05 01/2010

Strona 69

Lab 86071ba75554e05 01/2010Appendix: Firmware updateGeneral information With the Update_Labxxx_MxxxP program and a PC you can update the firmware of t

Strona 70

Lab 86072ba75554e05 01/2010After switching the meter off and on you can check whether the meter has taken over the new software version (see page 57).

Strona 72

SI Analytics GmbHPostfach 2443D-55014 MainzHattenbergstr. 10D-55122 MainzEmail: support@Telefon +49 (0) 61 31/66 5111Telefax +49 (0) 61 31/66 5001si-a

Strona 73

Overview Lab 8608ba75554e05 01/20101.1 KeypadIn this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <OK>) in

Strona 74

Lab 860 Overview9ba75554e05 01/20101.2 DisplayFunction displayindicators888881888°C°FU pH%LoBatAutoStoreSCTPCalErrorTimeDay.MonthYearBaudNo.IdentmV/pH

Powiązane modele: Lab 860 BENCHTOP pH METER

Komentarze do niniejszej Instrukcji

Brak uwag