Lab 860 Operation
37
ba75554e05 01/2010
Download to display via
calibration menu
While the calibration data is being displayed, you can:
z press <OK> to display further calibration data
(date, zero point, slope)
z press
<><>
to switch over the unit of the zero point (Asy)
(while the zero point is being displayed)
z press
<><>
to switch over the unit of the slope (Slo)
(while the slope is being displayed)
z press
<CAL
_>
to download the calibration record to the
interface
1 Press <CAL_> to display the calibration data.
The following data is displayed consecutively for 10 seconds
each:
date, zero point, slope.
z press <OK> to display further calibration data
(date, zero point, slope)
z press
<><>
to switch over the unit of the zero point (Asy)
(while the zero point is being displayed)
z press
<><>
to switch over the unit of the slope (Slo)
(while the slope is being displayed)
z press
<CAL
_>
to download the calibration record to the
interface