Xylem Lab 860 Instrukcja Użytkownika Strona 52

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 74
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 51
Operation Lab 860
52
ba75554e05 01/2010
4.8.2 Resetting all meter settings
This function resets to the default condition all meter settings including
the calibration values. The corresponding values are given on the
following pages:
Resetting the meter
settings
Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.
Calibration values page 51
System settings page 44
Measurement settings page 46
1 Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2 During the display test, press <MODE> to open the menu for
the reset of the meter settings.
Init is displayed.
3 Press <><> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
4 Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.
o
n
t
i
n
I
Przeglądanie stron 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 73 74

Komentarze do niniejszej Instrukcji

Brak uwag